FDA Class I (most serious) · Z-0316-2027 · Confirmed November 25, 2026
Is my Tandem Mobi insulin pump recalled?
Last updated: · Source: FDA consumer notice for Z-0316-2027
What was announced?
FDA has confirmed a most serious (Class I) recall of Tandem Mobi insulin pumps with Interoperable Technology (enforcement number Z-0316-2027). The consumer notice is dated November 25, 2026; Tandem sent letters on September 23, 2026. Certain pumps may issue false-positive occlusion alarms and suspend insulin until the user acknowledges the alarm and resumes therapy. If your serial number is on FDA’s official replacement list, request a return and replacement immediately at 877-801-6901 or the official Tandem RMA link. If it is not on that list, you may continue to use the pump while following the updated instructions (respond to every occlusion alarm, enable app notifications, and keep a backup insulin method as directed by your healthcare provider). As of October 28, 2026, Tandem reported 1,248 serious injuries and no deaths associated with this issue. This is not the November 12 Android Extended Bolus Class II recall (Z-0437-2027).
November 25, 2026 is the FDA consumer confirmation (“Content current as of: 11/25/2026”) and the “Recall Confirmed” index row — not the start date. The firm initiated the action on September 23, 2026; the Class 1 / Class I record was posted October 30, 2026; the OpenFDA report date is November 11, 2026. The recalling firm on the classification record is Tandem Diabetes Care, Inc., 12400 High Bluff Dr, San Diego, CA 92130-3077. The FDA consumer page shortens the name to Tandem Diabetes.
The official quantity in commerce is 80,938 (77,559 US, 3,379 OUS). That figure is the circulation total in the enforcement record. It is not the number of pumps that must be replaced, and it should not be added to the separate list of serials “to be exchanged.” The official status on the RES record is Open, Classified; OpenFDA lists the status as Ongoing. The action is voluntary and firm-initiated. Official notices do not list a sold-at window, retail channels, or price for this device recall, and this page does not invent them.
FDA’s opening characterization is: “This recall involves removing certain devices from where they are used or sold. This recall also involves updated use instructions for devices not subject to removal.” FDA has identified this recall as the most serious type. “This device may cause serious injury or death if you continue to use it without following the recommendations below.”
Which models are included?
The FDA consumer page says the action covers all Tandem Mobi Insulin Pumps with Interoperable Technology, with two different official recommendation tracks based on the serial number. The classified enforcement record lists these 20 model/catalog numbers, in official order (not strict numeric order): 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078.
This site also has an automatic FDA record page for Z-0316-2027. That archive slug names only the first five catalog numbers and is a record page, not a substitute for this consumer checklist. Model 1015472 is not in this 20-model list; it belongs to the separate Android Extended Bolus Class II action described below.
The official code-information field lists 9 UDI-DIs, in official order: 00389152075013, 00389152271910, 00389152350110, 00389152365510, 00389152365619, 00389152370019, 00389152480114, 00389152481111, 00389152481319. The enforcement record does not list a UDI-DI for the other 11 models, and this page does not invent those identifiers. The official consumer photo is labeled “Figure 1. Front and Back Images of the Tandem Mobi Insulin Pump”: FDA image of the Tandem Mobi insulin pump. The Tandem Mobi insulin pump is intended for subcutaneous delivery of insulin for diabetes management. The pump communicates with compatible devices and can be used with interoperable continuous glucose monitors and automated insulin dosing systems to automatically adjust insulin delivery.
Does my serial have to be replaced?
Check your pump’s serial number against FDA’s official List of Affected Product Subject to Replacement. Official sources fetched for this page do not say where the serial number is printed on the pump, and this page does not invent a label location or app menu path. Use the serial on the device or in your Tandem records and compare it with that official file.
The recommendations differ by serial. If your serial is on that replacement list, follow track 1: request return and replacement immediately. If it is not on the list, follow track 2: you may continue to use the pump while following the updated instructions. The official code-information qualifier also says: “All affected pump models and serial numbers may be affected.” Being off the replacement list does not mean the pump is outside the action.
The same official records include a list of “Pump serial numbers with frequent false-positive occlusion alarms to be exchanged.” Independently counted from the official RES / OpenFDA code-information text, that exchange list has 4,605 unique 7-digit serials. That count is not the recall total, is not a substitute for 80,938, and must not be added to 80,938. This page does not paste that serial list; use the official FDA attachment and the classification record.
What is the hazard?
Tandem Diabetes stated that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
The RES / OpenFDA manufacturer-reason text is consistent with that consumer-page reason and adds this investigation clause, quoted here as RES/OpenFDA wording (it does not appear on the November 25 consumer page itself): “Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold.”
What should I do?
Official recommendations depend on whether your serial is on the replacement list. This page does not invent a “stop using the pump immediately” order, and it does not describe a software update as the remedy for this occlusion-alarm action.
If your serial is on the replacement list
Confirm the pump on the official serial-number look-up list above. Then, in the FDA consumer notice’s words: “Immediately submit a request for return and replacement of your pump using the link provided in the customer letter from Tandem Diabetes or by contacting Tandem Customer Technical Support at 877-801-6901.” The official replacement request page linked from the FDA notice is campaign.tandemdiabetes.com/Mobi-Occlusions-RMA-Request.
Always respond to an occlusion alarm and follow the User Guide troubleshooting steps: acknowledge the alarm, inspect the cartridge, tubing, and infusion site, correct any identified condition, and resume insulin delivery. “If you continue to use the pump,” enable Tandem Mobi mobile app push notifications. The pump also notifies with audible tones or vibrations, progressing to louder tones until acknowledged. Regularly check blood sugar for unexpectedly high or low readings. Be prepared to use a backup method of insulin delivery “as directed by your healthcare provider and described in your pump training and User Guide.”
If your serial is not on the replacement list
The FDA consumer notice’s lead for all other Tandem Mobi users is: “You may continue to use your pump while following the instructions below.” Always respond to an occlusion alarm and follow the same User Guide troubleshooting steps. Enable Tandem Mobi mobile app push notifications so that an occlusion alarm is also delivered by the app; the pump still uses audible tones or vibrations, progressing to louder tones until acknowledged. If you “start to experience an unusually high number of occlusion alarms, or occlusion alarms unexpectedly and repeatedly disrupt your therapy even after following troubleshooting steps, contact Tandem Customer Support.” Be prepared to use a backup method of insulin delivery as directed by your healthcare provider and described in your pump training and User Guide.
The official track 2 list does not repeat the “regularly check your blood sugar” bullet that appears in track 1. This page does not add that bullet to track 2 as if it were printed there.
How do I request a replacement or get notification help?
U.S. customers should contact Tandem at [email protected] or 877-801-6901. The official recall notices and the RES Action field do not list support hours, and this page does not invent them. The RES / RMA materials write the same number as 1-877-801-6901.
The FDA notice also points to a Tandem support article for enabling smartphone notifications: Enable smartphone notifications to receive alerts. That official FDA-linked article’s body describes the Tandem t:slim mobile app and t:slim X2, including Android steps that say to make sure “Tandem t:slim is selected.” It is the official linked article; this page does not rewrite those screenshots or steps as if the article were already Mobi-specific UI copy.
The RES Action field also says that on 9/23/2026, “URGENT MEDICAL DEVICE CORRECTION” recall and correction notices were emailed, and that correction-notice customers were asked to complete an online acknowledgement form at campaign.tandemdiabetes.com/FCN-Mobi-Occlusions. The live campaign pages use the heading “URGENT MEDICAL DEVICE RECALL.” The acknowledgement form is a RES Action item for correction notices; it is not printed in the FDA consumer track 1 / track 2 lists. Healthcare providers and United States distributors and pharmacies also received letters. That statement is not a sold-at claim.
This page is not clinical advice beyond the official “as directed by your healthcare provider” / User Guide language. See what to do with a recalled product for general steps, and check other FDA food and drug / device recall records.
Were injuries or deaths reported?
The FDA consumer notice says, verbatim: “As of October 28, 2026, Tandem Diabetes has reported 1,248 serious injuries and no deaths associated with this issue.” October 28, 2026 is the official as-of date for that count. November 25, 2026 is the consumer-notice date, not the injury as-of date.
Is this the Android Extended Bolus recall, a t:slim recall, or the Malfunction 12 software correction?
No. Keep these actions separate, and do not add their unit or injury counts to this recall’s 80,938 quantity or to the 1,248 serious injuries reported as of October 28, 2026.
The November 12 Android Extended Bolus action is a separate Class II recall, Z-0437-2027 (event 100091). It covers Tandem Mobi with Interoperable Technology (with Android Tandem Mobi Mobile app), model/catalog 1015472, UDI-DI 00389152547244, Android software 3.5.4, 3.5.2.2, and 3.5.2.1, and about 4,086 users. The firm initiated that action on October 13, 2026, and it was posted on November 12, 2026. Its instruction is not to use Extended Bolus on the Android app; Standard Bolus and other functions may continue. That instruction does not apply to this occlusion-alarm recall, and this page’s replacement / updated-use instructions do not apply to Z-0437-2027. Official RES record: Z-0437-2027 classification record. This site’s automatic record page for that action is the Z-0437-2027 archive record, which is not a substitute for this landing.
The 2025 Malfunction 12 / false motor-failure correction is a different Class I action, Z-0427-2026, with a different hazard, a software-update remedy called version 7.9.0.2, and a different official injury sentence dated as of November 4, 2025. Do not treat “update to 7.9.0.2” as the fix for this false-occlusion-alarm recall. FDA consumer page: Insulin Pump Correction: Tandem Diabetes Care Issues Correction for Tandem Mobi Insulin Pumps. Optional automatic record: the Z-0427-2026 archive record.
A “Tandem recall” headline may also refer to a t:slim X2 pump, a t:connect or t:slim mobile-app action, or another Mobi campaign. Those are separate notices. This landing is only Z-0316-2027.
Official sources
- FDA consumer notice: Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps (content current as of 11/25/2026)
- FDA List of Affected Product Subject to Replacement (official serial look-up file)
- FDA RES classification record for Z-0316-2027 (Class 1, id=222734, event 99851)
- FDA RES event search for recall event 99851
- Tandem RMA / return-and-replacement request page (linked from the FDA consumer notice)
- FDA MedWatch: report an adverse reaction or quality problem
The automatic FDA record page for this enforcement number is the Z-0316-2027 archive record. It is a record page, not a substitute for this curated consumer checklist. Adverse reactions or quality problems can also be reported to FDA MedWatch or by calling FDA at 1-888-INFO-FDA (1-888-463-6332). Official distribution text from the enforcement record, including the official token “null,” is: Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada. Canada is the only named foreign country. The 3,379 OUS units are not broken out by additional countries in the official record.
Identification, dates, the two-track remedy, contacts, models, quantity in commerce, UDI-DIs, and the injury sentence come from the FDA consumer notice and the Z-0316-2027 classification / OpenFDA record. Company campaign pages are cited only where they are linked or labeled as such. Not affiliated with FDA or Tandem Diabetes Care, Inc.
Related: Check other FDA food and drug / device recall records · What to do with a recalled product · Latest recalls · Check any product
Recalled or Not? is an independent search aid. This page is not medical advice beyond the official FDA and RES wording cited above. Follow the linked official sources if the remedy, serial list, or contact details change.
FAQ
Is there a Tandem Mobi insulin pump recall in 2026?
Yes. On November 25, 2026, FDA posted a confirmed most serious recall of Tandem Mobi insulin pumps with Interoperable Technology because of false-positive occlusion alarms that can suspend insulin. The classified enforcement record is Z-0316-2027 (Class I), initiated by the firm on September 23, 2026 and posted October 30, 2026.
Which models are included?
The official enforcement list has 20 model/catalog numbers: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, and 1018078. The FDA consumer page also says the action covers all Tandem Mobi pumps with Interoperable Technology.
Does my pump have to be replaced, and how do I check the serial?
If your serial is on FDA’s “List of Affected Product Subject to Replacement” (https://www.fda.gov/media/197063/download?attachment), request return and replacement immediately (RMA link or 877-801-6901). If it is not on that list, you may continue to use the pump while following the updated instructions. Official records also say all affected models and serials may be affected.
What is the hazard?
Certain pumps may generate false-positive occlusion alarms when no insulin-flow obstruction is present, incorrectly suspend insulin, and leave delivery stopped until the user acknowledges the alarm and resumes therapy. Repeated interruptions can raise blood glucose; in severe cases hyperglycemia may need medical intervention or hospitalization.
Should I stop using the pump right now?
Follow the official track for your serial. Listed pumps: immediately request replacement; if you continue to use the pump, respond to every occlusion alarm, enable app notifications, check blood sugar, and have backup insulin as directed by your healthcare provider. Unlisted pumps: you may continue to use the pump under those updated instructions and contact Tandem if alarms become unusually frequent. This page does not invent a “stop immediately” order.
Were injuries or deaths reported?
As of October 28, 2026, Tandem Diabetes has reported 1,248 serious injuries and no deaths associated with this issue.
Is this the same as the November 12 Android Extended Bolus recall?
No. That is a separate Class II action, Z-0437-2027, for a software defect in the Android Tandem Mobi app (versions 3.5.4, 3.5.2.2, and 3.5.2.1), model 1015472, about 4,086 users. Its instruction is not to use Extended Bolus on Android. Do not mix the two, and do not add the unit counts.
How do I contact Tandem or report a problem?
U.S. customers should contact Tandem at [email protected] or 877-801-6901. Adverse reactions or quality problems can also be reported to FDA MedWatch. Official recall notices do not list support hours.
Update log
- Nov 26, 2026: Checked against the FDA consumer notice for the Tandem Mobi insulin pump recall, the official List of Affected Product Subject to Replacement, the RES / OpenFDA record for Z-0316-2027, and the live Tandem RMA page.