Tandem Mobi Insulin Pump Recall
Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936…
What is the FDA recall status?
The FDA record lists this recall as ongoing. FDA enforcement data is not a live recall-lifecycle tracker. Follow the official instructions and check the agency notice or firm for current guidance.
Why was Tandem Mobi Insulin Pump recalled?
Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
What should I do if I have this product?
Follow the firm and FDA instructions for the affected product. Contact the recalling firm for current guidance.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
The official FDA identifier list is too long to reproduce here. Compare your item with the complete list in the agency notice.
Full product description
Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078
Which lot codes, UPCs or models are affected?
The official FDA record has an unusually long identifier list. Check the complete lot, serial and model numbers in the linked agency notice.
Open the complete official FDA record
- Agency record number
- Z-0316-2027
- Product
- Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936…
- Recalling firm
- Tandem Diabetes Care, Inc.
- Quantity
- 80,938 (77,559 US, 3,379 OUS)
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada.
- Recall started
- September 23, 2026
- Report date
- November 11, 2026
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)