FDAClass IRecall started August 14, 2026

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories (MDR510094 PENDANT) Recall

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E…

What is the FDA recall status?

The FDA record lists this recall as ongoing. FDA enforcement data is not a live recall-lifecycle tracker. Follow the official instructions and check the agency notice or firm for current guidance.

How FDA enforcement data reports status

Why was Medline Basic Homecare Beds recalled?

Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.

What should I do if I have this product?

Follow the firm and FDA instructions for the affected product. Contact the recalling firm for current guidance.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare your product with the product description and available lot, UPC or model codes below from the official FDA record to check whether it is covered by this recall. If an identifier is missing or unclear, check the complete official notice or contact the firm; an incomplete match here does not establish that your product is unaffected.

Which lot codes, UPCs or models are affected?

MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots

Agency record number
Z-3227-2026
Product
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E…
Recalling firm
Medline Industries, LP
Quantity
771,127 total
Distribution
Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
Recall started
August 14, 2026
Report date
October 7, 2026
Classification
Class I
Status
Ongoing
Agency
FDA (device)