Operating Table TruSystem 7000 U (Catalog 1723633) Recall
Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
What is the FDA recall status?
The FDA record lists this recall as ongoing. FDA enforcement data is not a live recall-lifecycle tracker. Follow the official instructions and check the agency notice or firm for current guidance.
Why was Operating Table TruSystem 7000 U recalled?
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
What should I do if I have this product?
Follow the firm and FDA instructions for the affected product. Contact the recalling firm for current guidance.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare your product with the product description and available lot, UPC or model codes below from the official FDA record to check whether it is covered by this recall. If an identifier is missing or unclear, check the complete official notice or contact the firm; an incomplete match here does not establish that your product is unaffected.
Full product description
Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
Which lot codes, UPCs or models are affected?
Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 00887761967519
- Agency record number
- Z-0353-2027
- Product
- Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 62 units
- Distribution
- Worldwide - US Nationwide distribution.
- Recall started
- September 29, 2026
- Report date
- November 11, 2026
- Classification
- Class II
- Status
- Ongoing