FDAClass IIRecall started August 20, 2026

Fisher Scientific Sure-Vue Urine hCG Recall

Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;

What is the FDA recall status?

The FDA record lists this recall as ongoing. FDA enforcement data is not a live recall-lifecycle tracker. Follow the official instructions and check the agency notice or firm for current guidance.

How FDA enforcement data reports status

Why was Fisher Scientific Sure-Vue Urine hCG recalled?

hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.

What should I do if I have this product?

Follow the firm and FDA instructions for the affected product. Contact the recalling firm for current guidance.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare your product with the product description and available lot, UPC or model codes below from the official FDA record to check whether it is covered by this recall. If an identifier is missing or unclear, check the complete official notice or contact the firm; an incomplete match here does not establish that your product is unaffected.

Which lot codes, UPCs or models are affected?

All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900526/00613647000526/1155811105, 2018-01-15; 9004074/10304040024116/1155875704, 2014-06-09; 9004073/10304040024109/1155875704, 2014-06-09; 92217/10815845020109/1156015703, 07/16; B1077-22/20885380020251/1155904704, 9/2014; MPH12025HCG/10888277191877/1156093302, 2015-03-20; P080063/M220P0800633/40001853, 2023-03-30

Agency record number
Z-3274-2026
Product
Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;
Recalling firm
Alere San Diego, Inc.
Quantity
18,139,960
Distribution
Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.
Recall started
August 20, 2026
Report date
October 7, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls