FDAClass IAnnouncement date November 12, 2026

Luna G3 BPAP System LG3800-30VT — Class I recall for Targeted VT pressure-recording error

Luna G3 BPAP System for treatment of adult obstructive sleep apnea. Match the individual serial number against FDA Code Information; model alone does not establish inclusion.

Manually maintained record

An editor supplied or changed this record. These edits have not been verified by the agency. Last edited: 2026-11-20T00:57:13.978Z. The underlying agency record was last checked on 2026-11-25T10:29:08.313Z; that check does not verify the manual edits.

Reported risk

When Targeted VT is enabled, the configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure. This does not establish that the delivered pressure itself is incorrect. FDA classification: Class I; cause: device design. Incident and injury counts are not disclosed in this record.

What should I do if I have this product?

Patients using Targeted VT on the Luna G3 BPAP (LG3800-30VT): Contact your healthcare provider or durable medical equipment (DME) supplier immediately. Keep using your device as prescribed until your healthcare provider advises otherwise. Do not change your prescribed settings or stop respiratory therapy on your own. Contact your DME supplier to arrange the firmware correction.

Healthcare professionals and DME suppliers: Immediately disable Targeted VT if enabled on affected devices and coordinate continued therapy with an unaffected PAP device as directed by the treating clinician. Segregate unused affected units and remove affected units from clinical service until corrected. Interpret historical Targeted VT pressure reports with caution; they may not reflect delivered pressure. Coordinate firmware G3-2.00.80 with React Health. The update does not correct previously recorded data.

Source: React Health correction update, published by FDA on November 19, 2026.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

The product description and identifiers below were supplied or edited by a person. Confirm them with the agency notice or the firm before relying on a match; missing information does not establish that your product is unaffected.

Which lot codes, UPCs or models are affected?

UDI-DI 06948538365731. Individual affected serial numbers are listed in the official FDA notice. Recall Event 100001; Recall Z-0409-2027.

Agency record number
Z-0409-2027
Affected products
Luna G3 BPAP System for treatment of adult obstructive sleep apnea. Match the individual serial number against FDA Code Information; model alone does not establish inclusion.
Company
3B Medical, Inc.
Quantity
12,672 units
Distribution
United States nationwide including Puerto Rico; no distribution outside the United States.
Models
LG3800-30VT
Contact
FDA-listed company contact: George Koos, 1-863-2266285.
Report date
November 18, 2026
Classification
Class I
Status
Ongoing
Agency
FDA (device)

Notice details

FDA posted this Class I recall on November 12, 2026. The firm initiated the action on October 8, 2026 and began notifying customers on October 9, 2026. FDA status at review: Open, Classified. Official source checked November 13, 2026 UTC: FDA CDRH cfRES record 223102, Event 100001, Z-0409-2027. The official notice contains a long affected-serial-number list; consult that list and the notification letter to identify an individual device. The record does not disclose injury or death counts. A specific event revision timestamp is not disclosed.