VERIFY Biological Steam Test Pack (S3065) Recall
VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
What is the FDA recall status?
The FDA record lists this recall as ongoing. FDA enforcement data is not a live recall-lifecycle tracker. Follow the official instructions and check the agency notice or firm for current guidance.
Why was VERIFY Biological Steam Test Pack recalled?
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
What should I do if I have this product?
Follow the firm and FDA instructions for the affected product. Contact the recalling firm for current guidance.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare your product with the product description and available lot, UPC or model codes below from the official FDA record to check whether it is covered by this recall. If an identifier is missing or unclear, check the complete official notice or contact the firm; an incomplete match here does not establish that your product is unaffected.
Which lot codes, UPCs or models are affected?
UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
- Agency record number
- Z-0074-2027
- Product
- VERIFY Biological Steam Test Pack Model/Catalog Number: S3065
- Recalling firm
- Steris Corporation
- Quantity
- 217
- Distribution
- Worldwide - US Nationwide distribution.
- Recall started
- August 31, 2026
- Report date
- October 14, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)