FDAClass IIOctober 2, 2024

BD Phoenix 100-Intended for the rapid identification recall: E. coli

BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD…

Why was BD Phoenix 100-Intended for the rapid identification recalled?

Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI: 00382904481001 All serial numbers

Product
BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD…
Recalling firm
Becton Dickinson & Co.
Quantity
764 (EA)
Distribution
Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Dominican Republic Ecuador, Egypt, El Salvador, Ethiopia, France, French Guiana, Georgia, Germany, Ghana, Greece, Guatemala, Guinea, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Mali, Mauritania, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Syrian Arab, Republic Taiwan, Tanzania, United Republic of Thailand, Timor-Leste, Trinidad And Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, Vietnam, Zambia, Zimbabwe.
Recall date
October 2, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)