FDAClass IIOctober 4, 2017

Platinium DR DF4 1540 recall

Platinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator, Biventricular or ventricular…

Why was Platinium DR DF4 1540 recalled?

There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac ResynchronizationTherapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Serial Numbers

Product
Platinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator, Biventricular or ventricular…
Recalling firm
Sorin Group Italia SRL - CRF
Quantity
86 units
Distribution
US Nationwide Distribution
Recall date
October 4, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)