Bearing and Axle Bushing Assy STD Elbow C-m113-1-301 recall
BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800…
Why was Bearing and Axle Bushing Assy STD Elbow C-m113-1-301 recalled?
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
C-M113-1-301 UDI-DI 07613327590074 Lot MH621A C-M113-5-600 No UDI Lot 194593 C-M063-5-800 No UDI Lot RX06358A
- Product
- BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800…
- Recalling firm
- Howmedica Osteonics Corp.
- Quantity
- 12,156 US
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Recall date
- September 23, 2026
- Classification
- Class II
- Status
- Ongoing