FDAClass IISeptember 23, 2026

Captus 4000e Thyroid Uptake System recall

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Why was Captus 4000e Thyroid Uptake System recalled?

Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.

Product
Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;
Recalling firm
Mirion Technologies (Capintec), Inc.
Quantity
783
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
Recall date
September 23, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)