ECT Internal Fracture Fixation Quick-Connect Drill 2.5 recall
ECT Internal Fracture Fixation Quick-Connect Drill 2.5 mm Diameter 12.5 cm Length, Item Number/EDI 00241002600, Nonsterile. orthopedic surgical…
Why was ECT Internal Fracture Fixation Quick-Connect Drill 2.5 recalled?
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: 63289918, 63296044, 63451593, 63457958, 63462351, 63469945, 63479047 & 63518080.
- Product
- ECT Internal Fracture Fixation Quick-Connect Drill 2.5 mm Diameter 12.5 cm Length, Item Number/EDI 00241002600, Nonsterile. orthopedic surgical…
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 198 units
- Distribution
- Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
- Recall date
- October 4, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)