FDAClass IISeptember 26, 2018

PR1000 Pivot Rehab Bed AC powered adjustable bed recall

PR1000 Pivot Rehab Bed AC powered adjustable bed.

Why was PR1000 Pivot Rehab Bed AC powered adjustable bed recalled?

Under certain conditions, when pivoting bed to the extreme left there are situations where the wire/cables may become pinched. A failure would result in the resident requiring to be transferred to a different bed until repairs can be made

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ALL Pivot Rehab PR1000 Beds built between 12/01/2010 thru 12/22/2014 SN: PR 10001 build date 12/01/2010 thru PR 10255 build date 12/22/2014

Product
PR1000 Pivot Rehab Bed AC powered adjustable bed.
Recalling firm
Med-Mizer, Inc.
Quantity
230 units
Distribution
The products were distributed to the following US states: AR, AZ, FL, GA, IA, ID, IL, IN, KY, ME, MI, MN, MO, NC, ND, NE, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, and WI. The products were distributed to the following foreign countries: Canada, China, and Denmark.
Recall date
September 26, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls