FDAClass IISeptember 25, 2024

System Name: 722044 INTEGRIS Allura 15-12 recall

System Name: 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEGRIS Allura 9 (biplane) 722018 Integris…

Why was System Name: 722044 INTEGRIS Allura 15-12 recalled?

Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Product UDI-DI: N/A 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEGRIS Allura 9 (biplane) 722018 Integris Allura 9 722017 Integris H5000 F / Allura 9F 722016 Integris H5000 C / Allura 9C All serial numbers

Product
System Name: 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEGRIS Allura 9 (biplane) 722018 Integris…
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
250 units
Distribution
Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen
Recall date
September 25, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)