FDAClass IISeptember 25, 2024

System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 recall

System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228…

Why was System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 recalled?

Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model Product UDI-DI: 1. 722282 Azurion 7 M20 (01)00884838110571(21); 2. 722281 Azurion 5 M20 (01)00884838110564(21); 3. 722236 Azurion 7 B20 (01)00884838116801(21); 4. 722235 Azurion 7 B12 (01)00884838116788(21); 5. 722234 Azurion 7 M20 (01)00884838116771(21); 6. 722233 Azurion 7 M12 (01)00884838116764(21); 7. 722228 Azurion 5 M20 (01)00884838099234(21); 8. 722227 Azurion 5 M12 (01)00884838099227(21); 9. 722226 Azurion 7 B20 (01)00884838099272(21); 10. 722225 Azurion 7 B12 (01)00884838099265(21); 11. 722224 Azurion 7 M20 (01)00884838099258(21); 12. 722223 Azurion 7 M12 (01)00884838099241(21); 13. 722222 Azurion 3 M15 (01)00884838099210(21); 14. 722221 Azurion 3 M12 (01)00884838099203(21); 15. 722079 Azurion 7 M20 (01)00884838085268(21); 16. 722078 Azurion 7 M12 (01)00884838085251(21); 17. 722068 Azurion 7 B20 (01)00884838085367(21); 18.722067 Azurion 7 B12 (01)00884838085350(21); 19. 722064 Azurion 3 M15 (01)00884838085282(21); 20. 722063 Azurion 3 M12 (01)00884838085275(21) All serial numbers

Product
System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228…
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
7091 units
Distribution
Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen
Recall date
September 25, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)