FDAClass IISeptember 25, 2024

Marodyne-LiV recall

Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

Why was Marodyne-LiV recalled?

Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

serial numbers: 010426065891001421S05327911240506, 010426065891001421S05328011240506, 010426065891001421S05328111240506, 010426065891001421S05328211240506, 010426065891001421S05328311240506, 010426065891001421S05328411240506, 010426065891001421S05328511240506, 010426065891001421S05328611240506, 010426065891001421S05328711240506, 010426065891001421S05328811240506, 010426065891001421S05328911240506, 010426065891001421S05329011240506, 010426065891001421S05321911240430, 010426065891001421S05322011240430, 010426065891001421S05322111240430, 010426065891001421S05322211240430, 010426065891001421S05322311240430, 010426065891001421S05322411240430, 010426065891001421S05322511240430, 010426065891001421S05322611240430, 010426065891000721S05336411240521, 010426065891000721S05336511240521, 010426065891000721S05336611240521, 010426065891000721S05336711240521, 010426065891000721S05336811240521, 010426065891000721S05336911240521, 010426065891000721S05337011240521, 010426065891000721S05337111240521, 010426065891000721S05337211240521, 010426065891000721S05337311240521, 010426065891000721S05337411240521, 010426065891000721S05337511240521, 010426065891000721S05337611240521, 010426065891000721S05337711240521, 010426065891000721S05337811240521, 010426065891000721S05337911240521, 010426065891000721S05338011240521, 010426065891000721S05338111240521, 010426065891000721S05338211240521, 010426065891000721S05338311240521, 010426065891000721S05338411240521, 010426065891000721S05338511240521, 010426065891000721S05338611240521, 010426065891000721S05338711240521, 010426065891000721S05338811240521, 010426065891000721S05338911240521, 010426065891000721S05339011240521, 010426065891000721S05339111240521, 010426065891000721S05339211240521, 010426065891000721S05339311240521, 010426065891000721S05339411240521, 010426065891000721S05339511240521, 010426065891000721S05339611240521, 010426065891000721S05339711240521, 010426065891000721S05339811240521, 010426065891000721S05339911240521, 010426065891000721S05340011240521, 010426065891000721S05340111240521, 010426065891000721S05340211240521, 010426065891000721S05340311240521, 010426065891000721S05340411240521, 010426065891000721S05340511240521

Product
Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment
Recalling firm
Btt Health Gmbh
Quantity
62 units
Distribution
US Nationwide distribution in the states of Florida.
Recall date
September 25, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)