The device is a mobile x-ray system designed to work recall
The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray…
Why was The device is a mobile x-ray system designed to work recalled?
When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All serial numbers shipped from FUJIFILM
- Product
- The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray…
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Quantity
- 130
- Distribution
- US Nationwide and Worldwide
- Recall date
- October 2, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)