IV Non-Dedicated Gravity Administration Sets with drip recall
IV Non-Dedicated Gravity Administration Sets with drip chambers
Why was IV Non-Dedicated Gravity Administration Sets with drip recalled?
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All products manufactured on or before 7/4/2026. UDI-DI: 10887709079363 10887709079370 10887709079417 10887709079431 10887709079448 10887709079455 10887709097473 10887709079493 10887709079509 10887709079547
- Product
- IV Non-Dedicated Gravity Administration Sets with drip chambers
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 2,450
- Distribution
- Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.
- Recall date
- September 23, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)