Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K recall
Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United…
Why was Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K recalled?
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lots
- Product
- Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United…
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 320 units
- Distribution
- Distribution US nationwide.
- Recall date
- September 20, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)