MOSAIQ Medical Charged-Particle Radiation Therapy recall
MOSAIQ Medical Charged-Particle Radiation Therapy System
Why was MOSAIQ Medical Charged-Particle Radiation Therapy recalled?
The table shift in the vertical direction was inverted during the use of CMA (Couch Move Assistant). Therapy may be directed to an incorrect area.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All software builds.
- Product
- MOSAIQ Medical Charged-Particle Radiation Therapy System
- Recalling firm
- Elekta, Inc.
- Quantity
- 2,461
- Distribution
- Worldwide distribution. US nationwide to AK, AL, AR, AZ, CA, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, District of Columbia, Hawaii and Puerto Rico. ALBANIA, ALGERIA, ANTIGUA AND BARBUDA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BENIN, BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CUBA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, FINLAND, FRANCE, GABON, GEORGIA, GERMANY, GHANA, GREECE, HONDURRAS, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, LAOS, LIBYA, LITHUANIA, MALAYSIA, MALTA, MEXICO, MOROCCO, MYANMAR, NAMIBIA, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, VENEZUELA, VIETNAM and ZIMBABWE
- Recall date
- September 13, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)