Artis zee/zeego and Artis Q/Q.zen system recall
Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for…
Why was Artis zee/zeego and Artis Q/Q.zen system recalled?
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can results in a cooling system failure and subsequent system failure. This leek can occur sporadically in affected systems.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Artis Q floor - Serial number: 103030 103013 103039 103008 103035 103001 103028 103040
- Product
- Artis zee/zeego and Artis Q/Q.zen system, x-ray, angiographic AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 986 units distributed in the USA
- Distribution
- Nationwide Distribution
- Recall date
- September 6, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)