FDAClass IISeptember 6, 2017

Polyurethane Enfit(r) Ameritus NG Feeding Tube recall

POLYURETHANE ENFit(R) AMERITUS NG FEEDING TUBE, WITH SIDE PORT, REF Numbers: a) ENF-Y50P-40 b) ENF-Y50P-50 c) ENF-Y60P-40 d) ENF-Y60P-50 e)…

Why was Polyurethane Enfit(r) Ameritus NG Feeding Tube recalled?

Inadequate detectability of radiopaque stripes

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

a) ENF-Y50P-40, lot number KS1608011 b) ENF-Y50P-50, lot number KS1608012 c) ENF-Y60P-40, lot number KS1608013 d) ENF-Y60P-50, lot number KS1608014 e) ENF-Y60P-90, lot number KS1608015 f) ENF-Y80P-40, lot number KS1606035 g) ENF-Y80P-50, lot number KS1606036 h) ENF-Y80P-80, lot number KS1606037 i) ENF-Y80P-90, lot number KS1606038

Product
POLYURETHANE ENFit(R) AMERITUS NG FEEDING TUBE, WITH SIDE PORT, REF Numbers: a) ENF-Y50P-40 b) ENF-Y50P-50 c) ENF-Y60P-40 d) ENF-Y60P-50 e)…
Recalling firm
Kentec Medical Inc
Quantity
K100526
Distribution
TX, VA, IL, CA, KS
Recall date
September 6, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)