FDAClass IIOctober 7, 2020

Ac/dc Power Supply for 7400 recall

AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable…

Why was Ac/dc Power Supply for 7400 recalled?

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: Below K17510001

Product
AC/DC POWER SUPPLY for 7400, Part Number 151004300, XP Power Model Number AHM180PS19-XE0338, AC/DC Adapter associated with the following portable…
Recalling firm
ESAOTE S.P.A
Quantity
989 total adapters in the US
Distribution
US Nationwide
Recall date
October 7, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)