ProScreen" Drugs of Abuse Test recall: Labeling Error
ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC"…
Why was ProScreen" Drugs of Abuse Test recalled?
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
ALL Lots Within Expiry
- Product
- ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC"…
- Recalling firm
- Ameditech Inc
- Quantity
- 51,075 units
- Distribution
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
- Recall date
- October 7, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)