FDAClass IIOctober 7, 2020

ProScreen" Drugs of Abuse Test recall: Labeling Error

ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC"…

Why was ProScreen" Drugs of Abuse Test recalled?

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

ALL Lots Within Expiry

Product
ProScreen" Drugs of Abuse Test (AMT Item/ AMT Description): PSD-6MB/ "ProScreen" Drugs of Abuse Test (MET1000/OPI2000)+AMP+BZO+COC+THC"…
Recalling firm
Ameditech Inc
Quantity
51,075 units
Distribution
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Recall date
October 7, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls