FDAClass IISeptember 19, 2018

Vascular Solutions Venture recall: Labeling Error

Vascular Solutions Venture Rx Catheter, 6F, resterilized.

Why was Vascular Solutions Venture recalled?

The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Code Number 5820-R; Lot Number 570183 (Serial Number 99804132016-07); Lot Number 583022 (Serial Number 17304202017-07, 17304202017-08)

Product
Vascular Solutions Venture Rx Catheter, 6F, resterilized.
Recalling firm
SPS Sterilization, Inc
Quantity
3 units total
Distribution
US distribution to AZ, CA, FL, OH, PA, TX, VA, WV
Recall date
September 19, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls