Latest recalls
FDAClass IISeptember 2, 2026

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Why it was recalled

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

What to do

The recall is ongoing. Follow the firm and FDA instructions.

Affected products and identifiers

114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376 — 13 units

Record details

IMRIS Imaging Inc — Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.

Open FDA enforcement record