1.5T Signa HDx recall
1.5T Signa HDx, 3.0T Signa HDx, 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The GE Signa¿ HDx MR system is a whole body…
Why was 1.5T Signa HDx recalled?
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Medical Device Listing No. D117354, D295466, D182389, D102432, D206140, D230310, D269569, D102430, D140572, D188736, D102434, D234841, D196247, D134019, D236543, D006275, D159181, D288890, D229859; Model No. Eaton 9130, 5323286, 5396502-5, 5499460-3, 5499460-6, 5499460-9, 5741000-3, 00840682115971, 5396502-2, 5499460-2, 5741000-2, 00840682115872, 5373011-3, 5396469-3, 5481490-2, 5481500-7, 5499460-7, 5499460-10, 5741000, 00840682103817, 5323286-101, 2377062-61, 2395001, 2395001-2, 2395001-3, 2395001-4, 00840682105699, 840682105378
- Product
- 1.5T Signa HDx, 3.0T Signa HDx, 1.5T Signa HDxt, 3.0T Signa HDxt, Signa HDi, Signa Vibrant Product Usage: The GE Signa¿ HDx MR system is a whole body…
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 68 units in total
- Distribution
- US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
- Recall date
- August 23, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)