Medline breathing circuits labeled as: 1) MEDLINE recall
Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex…
Why was Medline breathing circuits labeled as: 1) MEDLINE recalled?
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
1) Medline Product Number/SKU HUD78001KIT (Teleflex Product Number/SKU 780-01KIT), UDI/DI 10197344010752 (each) 20197344010759 (case), All lot numbers; 2) Medline Product Number/SKU HUD78007KIT (Teleflex Product Number/SKU 780-07KIT), UDI/DI 10197344010738 (each) 20197344010735 (case), All lot numbers; 3) Medline Product Number/SKU HUD78008KIT (Teleflex Product Number/SKU 780-08KIT), UDI/DI 10197344010721 (each) 20197344010728 (case), All lot numbers; 4) Medline Product Number/SKU HUD78009KIT (Teleflex Product Number/SKU 790-09KIT), UDI/DI 10197344010714 (each) 20197344010711 (case), All lot numbers; 5) Medline Product Number/SKU HUD78017KIT (Teleflex Product Number/SKU 780-17KIT), UDI/DI 10197344010677 (each) 20197344010674 (case), All lot numbers; 6) Medline Product Number/SKU HUD78018KIT (Teleflex Product Number/SKU 780-18KIT), UDI/DI 10197344010905 (each) 20197344010902 (case), All lot numbers; 7) Medline Product Number/SKU HUD88001KIT (Teleflex Product Number/SKU 880-01KIT), UDI/DI 10197344010929 (each) 20197344010926 (case), All lot numbers; 8) Medline Product Number/SKU HUD88013KIT (Teleflex Product Number/SKU 880-13KIT), UDI/DI 10197344010493 (each) 20197344010490 (case), All lot numbers; 9) Medline Product Number/SKU HUD88014KIT (Teleflex Product Number/SKU 880-14KIT), UDI/DI 10197344010486 (each) 20197344010483 (case), All lot numbers; 10) Medline Product Number/SKU HUD88015KIT (Teleflex Product Number/SKU 880-15KIT), UDI/DI 10197344010479 (each) 20197344010476 (case), All lot numbers; 11) Medline Product Number/SKU HUD88017KIT (Teleflex Product Number/SKU 880-17KIT), UDI/DI 10197344010462 (each) 20197344010469 (case), All lot numbers;
- Product
- Medline breathing circuits labeled as: 1) MEDLINE HUDSON RCI, CIRCUIT, NEO, SING LIMB, HTD,COLUMN, Medline Product Number/SKU HUD78001KIT (Teleflex…
- Recalling firm
- Medline Industries, LP
- Quantity
- 271406 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Panama.
- Recall date
- September 2, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)