Latest recalls
FDAClass IISeptember 2, 2026

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing…

Why it was recalled

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

What to do

The recall is ongoing. Follow the firm and FDA instructions.

Affected products and identifiers

All lot codes — 20 units

Record details

Medline Industries, LP — Worldwide distribution - US Nationwide and the country of Panama.

Open FDA enforcement record