FDAClass IIAugust 26, 2026

Fort Defiance Industries recall

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Why was Fort Defiance Industries recalled?

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Code: All serial numbers/ GTIN: (01) 00862992000327

Product
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Recalling firm
Fort Defiance Industries, LLC
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Recall date
August 26, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)