FDAClass IISeptember 12, 2018

Laser CDS recall: Sterility Concern

LASER CDS

Why was Laser CDS recalled?

Multiple surgical packs were assembled with a non-sterile CHG applicators labeled as "ChloraPrep" (kit component numbers: 29183, 52115, 82279, and 82281). The non-sterile components 29183, 52115, 82279, and 82281 should have been placed inside the sterile portion of the pack, where it would subsequently be exposed to the ethylene oxide sterilization process. However, the components were not placed in the sterile portion of the pack during assembly and were never exposed to the sterilization process. Components 29183, 52115, 82279, and 82281 were inadvertently released non-sterile. The remainder of the pack components were properly sterilized and only components 29183, 52115, 82279, and 82281 were affected. Components 29183, 52115, 82279, and 82281 are manufactured by Becton, Dickinson (BD) and are labeled "ChloraPrep". ChloraPrep" applicators contain 2% w/v chlorhexidine gluconate (CHG) in 70% v/v isopropyl alcohol (IPA). The applicators are used to facilitate skin preparation for various preoperative and pre-injection procedures.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model: CDS980970D; Lots: 16FD2090, 16GD2340, 16ID0038, 16KD1940, 17BD1440, 17CD2208

Product
LASER CDS
Recalling firm
Medline Industries, Inc.
Quantity
144 kits
Distribution
CA, NM
Recall date
September 12, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls