FDAClass IISeptember 12, 2018

Laparoscopy CDS recall: Sterility Concern

Laparoscopy CDS

Why was Laparoscopy CDS recalled?

One manufacturing lot of a non-sterile pack was assembled with a non-sterile component labeled sterile (kit component #80532NS). The non-sterile component 80532NS was supposed to be placed inside the sterile portion of the pack where it would subsequently be exposed to the ethylene oxide sterilization process. However, the component was not placed in the sterile portion of the pack during assembly and was never exposed to the sterilization process. Component 80532NS was inadvertently released non-sterile. Component 80532NS is packaged by Medline and is an Anti-fog kit that contains 1 bottle of 6g Anti-Fog solution and an adhesive backed sponge.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model: CDS920114F. Lot: 17MB5499

Product
Laparoscopy CDS
Recalling firm
Medline Industries, Inc.
Quantity
15 kits
Distribution
CA, NM
Recall date
September 12, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)