FDAClass IISeptember 23, 2020

Imactis Patient Fiducial recall

Imactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the ImactisCT-Navigation Workstation (REF…

Why was Imactis Patient Fiducial recalled?

The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Serial Numbers

Product
Imactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the ImactisCT-Navigation Workstation (REF…
Recalling firm
Imactis
Quantity
6 Navigation Systems
Distribution
US Nationwide distribution including in the states of FL & WI.
Recall date
September 23, 2020
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls