TD Navio Maxi Speech Generating Device recall
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
Why was TD Navio Maxi Speech Generating Device recalled?
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
- Product
- TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
- Recalling firm
- Tobii Dynavox Llc
- Distribution
- US Nationwide distribution.
- Recall date
- August 26, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)