FDAClass IIAugust 26, 2026

TD Navio Midi Speech Generating Device recall

TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Why was TD Navio Midi Speech Generating Device recalled?

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).

Product
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Recalling firm
Tobii Dynavox Llc
Distribution
US Nationwide distribution.
Recall date
August 26, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls