FDAClass IISeptember 23, 2020

NovaPACS Diagnostic Viewer. NovaPACS is a picture recall

NovaPACS Diagnostic Viewer. NovaPACS is a picture archiving and communication system software that retrieves, distributes, and display images and data…

Why was NovaPACS Diagnostic Viewer. NovaPACS is a picture recalled?

The firm received a report of an atypical dataset being generated. When using the cross-localization feature and images from a modality that generates asymmetrical images, the cross-localization reference line may be inaccurately placed on any of the corresponding images that are open.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

NovaPACS versions 8.6.8, 8.6.9, 8.6.11, 8.6.12, 8.6.14, 8.7.4, 8.7.6, and 8.7.8

Product
NovaPACS Diagnostic Viewer. NovaPACS is a picture archiving and communication system software that retrieves, distributes, and display images and data…
Recalling firm
Novarad Corporation
Quantity
197 units
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Colombia, Dominican Republic, Indonesia, Jordan, Mexico, Peru, Phillipines, and United Kingdom.
Recall date
September 23, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls