FDAClass IISeptember 11, 2024

BD Pyxis QFill Replenishment Station recall: Software Issue

BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet system

Why was BD Pyxis QFill Replenishment Station recalled?

Due to software issue, there is a potential when a restock label is printed for a medication stored in a non-CUBIE location, that the incorrect bin for the medication is printed on the label.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

serial numbers: 44210411, 43880801, 44227426/ UDI: None

Product
BD Pyxis QFill Replenishment Station, REF: 138904-01, medication cabinet system
Recalling firm
CareFusion 303, Inc.
Quantity
3 units
Distribution
US Nationwide distribution in the states of CA, FL, KS, MN, NJ, NY, PA.
Recall date
September 11, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls