FDAClass ISeptember 16, 2020

Alaris System PC Unit recall

Alaris System PC Unit Model 8015

Why was Alaris System PC Unit recalled?

The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All serial numbers manufactured from April 07, 2017 to Present.

Product
Alaris System PC Unit Model 8015
Recalling firm
CareFusion 303, Inc.
Quantity
313740
Distribution
U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA
Recall date
September 16, 2020
Classification
Class I
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls