GEM Premier 7000 with iQM3 recall
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier…
Why was GEM Premier 7000 with iQM3 recalled?
The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part No.00077407508 UDI-DI 00195226000764 Lot 260504E; Part No. 00077407510 UDI-DI 00195226000665 Lot 260505B; Part No. 00077415004 UDI-DI 00195226000856 Lot 260505A; Part No. 00077415008 UDI-DI 00195226000757 Lot No.260505A; Part No. 00077415010 UDI-DI 00195226000658 Lot 260505C; Part No. 00077430010 UDI-DI 00195226000641 Lot 260505B
- Product
- GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier…
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 32 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
- Recall date
- August 12, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)