FDAClass IIOctober 5, 2016

Alaris Syringe Module recall: Software Issue

Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers…

Why was Alaris Syringe Module recalled?

A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software version 9.15

Product
Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers…
Recalling firm
CareFusion 303, Inc.
Quantity
12,000 units
Distribution
Worldwide Distribution: US(nationwide) and countries of: Canada, Australia and United Arab Emirates.
Recall date
October 5, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)