FDAClass IIAugust 29, 2018

DigitalDiagnost recall

DigitalDiagnost, Dual-Detector (Stitching Patient Support) 712052

Why was DigitalDiagnost recalled?

The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 363679 404164/SN0802462 378830/SN0802004 341693 358494 359442 396571/SN0802311 359449 346075 382021/SN0802313 374374 404298/SN0802474 402570/SN0802420 373501 373500 382272/SN0802113 368428 382012/SN0702520 376667 356957 392502/SN0802179 407423/SN0802638 359767 368619 358641 357697 381998/SN0702488 413829/SN0802631 358320 400364/SN0802522 404166/SN0802441 356981 390272/SN0802364 390273/SN0802366 412482/SN0802683 348997 396576/SN0802233 372848 373508 383084/SN0802127

Product
DigitalDiagnost, Dual-Detector (Stitching Patient Support) 712052
Recalling firm
Philips Medical Systems Gmbh, DMC
Distribution
US Nationwide; International to 66 countries
Recall date
August 29, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls