FDAClass IIAugust 12, 2026

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair recall: Software Issue

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

Why was munevo DRIVE and munevo DRIVE r-net Powered Wheelchair recalled?

A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699

Product
munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
Recalling firm
MUNEVO GMBH
Quantity
19
Distribution
Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.
Recall date
August 12, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls