Why was Ferno iNX Wheeled Stretcher recalled?

Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial #s: 15F001703, 15F001704, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001715, 15F001717, 15F001719, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001728, 15F001729, 15F001730, 15F001731, 15F001732, 15F001733, 15F001734, 15F001735, 15F001736, 15F001737, 15F001738, 15F001739, 15F001740, 15F001741, 15F001742, 15F001744, 15F001745, 15F001746, 15F001747, 15F001748, 15F001749, 15F001750, 15F001751, 15F001752, 15F001753, 15F001754, 15F001755, 15F001756, 15F001757, 15F001758, 15F001759, 15F001760, 15F001761, 15F001762, 15F001763, 15F001764, 15F001765, 15F001766, 15F001767, 15F001768, 15F001769, , 5F001770, 15F001771, 15F001772, 15F001773, 15F001774, 15F001775, 15F001776, 15F001777, 15F001778, 15F001779, 15F001780, 15F001781, 15F001782, 15F001783, 15F001784, 15F001785, 15F001786, 15F001787, 15F001788, 15F001789, 15F001790, 15F001791, 15F001793, 15F001794, 15F001795, 15F001796, 15F001797, 15F001798, 15F001804, 15F001805, 15F001806, 15F001807, 15F001808, 15F001809, 15F001810, 15F001811, 15F001812, 15F001813, 15F001814, 15F001815, 15F001816 & 15F001850.

Product
Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.
Recalling firm
Ferno-Washington Inc
Quantity
72 units
Distribution
Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.