FDAClass IIOctober 7, 2015

Zimmer Guide Wires for orthopedic procedures recall: Labeling Error

Zimmer Guide Wires for orthopedic procedures.

Why was Zimmer Guide Wires for orthopedic procedures recalled?

Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

00-2228-024-00 2.4MM Diameter Bullet Tip Guide Wire; 00-2255-025-00 Humeral Smooth Guide Wire 2.4mm 70cm Length; 00-2255-026-00 Humeral Bullet Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2237-033-00 Smooth Guide Wire 2.4mm Diameter 100cm Length; 47-2237-037-00 3.0MM Diameter Smooth Guide Wire 100cm Length; 47-2237-038-00 Bullet Tip Guide Wire 3.0mm Diameter 100cm Length; 47-2255-008-00 Ball Tip Guide Wire 2.4mm Diameter 70cm Length; 47-2255-008-01 Ball Tip Guide Wire 3.0mm Diameter 100cm Length All non-expired, distributed lots with a 10 year expiration date on the label.

Product
Zimmer Guide Wires for orthopedic procedures.
Recalling firm
Zimmer, Inc.
Quantity
112,568 units
Distribution
Worldwide Distribution - US (nationwide) including Puerto Rico and Virgin Islands., and to the countries of Canada, Brazil, Algeria, Mexico, Austria, Taiwan, Belgium, Hong Kong, Croatia, Singapore, Denmark, Bahamas, Egypt, Panama, Finland, Dominican Republic, France, Guatemala, Germany, Australia, Ireland, Japan,Italy, Haiti, Jordan, Malaysia, Kenya, India, Lebanon, Luxembourg, Malta, Mauritius, Morocco, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Africa, Spain, Sweden, Switzerland, United Kingdom and United Arab Emirates.
Recall date
October 7, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls