MiniMed Go App recall
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
Why was MiniMed Go App recalled?
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0
- Product
- MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
- Recalling firm
- Medtronic MiniMed, Inc.
- Quantity
- 3,465
- Distribution
- Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
- Recall date
- July 29, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)