FDAClass IISeptember 23, 2015

Angiographic Injector. Medrad Mark 7 Arterion Injection recall: Software Issue

Angiographic Injector. Medrad Mark 7 Arterion Injection System, which utilizes Software Version SW 005.006_SH; to be used specifically for the…

Why was Angiographic Injector. Medrad Mark 7 Arterion Injection recalled?

Software Version SW 005.006_SH, resulted in the removal of purge enforcement from traditional New-Case, Power Up and Syringe Change use cases while the injector head is in the upright position.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Catalog numbers: ART 700 PEDL, ART 700 TABL, ART 700 OCS, which utilize Software Version SW 005.006_SH. Affected serial numbers: 20329, 20570, 101148, 20392, 20425, 20427, 30046, 100945, 100186, 21255, 30086, 20025, 101153, 101159, 20714, 100873, 20437, 100033, 100249, 20333, 100511, 101206, 100002, 100890, 101208, 20323, 20914, 20342, 10172, 10074, 20195, 20196, 20850, 20929, 101106, 30020, 30024, 30033, 20252, 10087, 20851, 100009, 100864, 100335, 10095, 20302, 20306, 100013, 20918, 100141, 20158, 100784, 20735, 20422, 101149, 20842, 20018, 101114, 100056, 100075, 20434, 21024, 20311, 20763, 10055, 20786, 20701, 20729, 20040, 100092, 20501

Product
Angiographic Injector. Medrad Mark 7 Arterion Injection System, which utilizes Software Version SW 005.006_SH; to be used specifically for the…
Recalling firm
Bayer Healthcare
Quantity
71
Distribution
Worldwide Distribution. US Nationwide, Germany, France, UK, Netherlands, Sweden, and Canada.
Recall date
September 23, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls