FDAClass IIJuly 29, 2026

3mensio Workstation recall: Software Issue

3mensio Workstation (Vascular Fenestrated) software

Why was 3mensio Workstation recalled?

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5

Product
3mensio Workstation (Vascular Fenestrated) software
Recalling firm
PIE Medical Imaging B.V.
Quantity
165
Distribution
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
Recall date
July 29, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)