Olympus Scientific Solutions Americas Corporation recall
Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence
Why was Olympus Scientific Solutions Americas Corporation recalled?
Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Models : VCR and VMR.
- Product
- Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence
- Recalling firm
- Olympus Scientific Solutions Americas
- Quantity
- 146 units
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution
- Recall date
- August 9, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)