FDAClass IIAugust 2, 2017

Medtronic Visualase(R) Cooled Laser Applicator System recall

Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling…

Why was Medtronic Visualase(R) Cooled Laser Applicator System recalled?

Medtronic has become aware that the VCLAS 15MM TIP cap, also referred to as Tuohy Borst Adapter (TBA), may not be capable of creating a seal around the optical fiber to ensure adequate saline flow through the cooling system. When the TBA is unable to seal around the optical fiber, low retention force to the optical fiber may occur, causing saline to leak out of the TBA.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 211158059, 211560973, 211959879, 12473141, 212707035, 211158151, 211594328, 211977586, 212503634, 212745660, 211158152, 211599898, 211987873, 212504463, 212796559, 211181212, 211628257, 212127654, 212549456, 212912501, 211181547,211908695, 212141557, 212553428, 211184941, 211916062, 212148368, 212632884,211555011, 211926889, 212434500, 212647830, 211555017, 211950736, 212438405, 212701679

Product
Medtronic Visualase(R) Cooled Laser Applicator System (VCLAS) 15MM TIP, REF 9735561, (1) Laser Diffusing Fiber - 600um Core, 15mm Tip, (1) Cooling…
Recalling firm
Medtronic Navigation, Inc.
Quantity
214 units
Distribution
US Distribution to the states of : CA, GA, MN, FL, NY, IL, TX
Recall date
August 2, 2017
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls