FDAClass IISeptember 26, 2018

Temperature sensor Foley catheter 18FR recall

Temperature sensor Foley catheter 18FR, Catalogue Number 102205101880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter…

Why was Temperature sensor Foley catheter 18FR recalled?

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

S17005574

Product
Temperature sensor Foley catheter 18FR, Catalogue Number 102205101880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter…
Recalling firm
Degania Silicone, Ltd.
Quantity
153
Distribution
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Recall date
September 26, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls